Double-step calculation of the intraocular lens power in lens opacities excluding the measurement of axial eye length by optical biometry. A clinical case
https://doi.org/10.21516/2072-0076-2023-16-4-127-132
Abstract
The main source of intraocular lens (IOL) power calculation errors in phacoemulsification (PE) is axial length (AL) measurement inaccuracy in cases if optical biometry cannot be used.
bto describe an original double-step method of IOL power calculation in lens opacities which exclude AL measurement with optical biometry.
Material and methods. Patient S. (male, 71) with a bilateral immature cataract underwent right eye phacoemulsification (PE) with no IOL implantation. After that, AL was measured in the aphakic state with transparent optical media and used in combination with preoperative anterior segment biometry data to calculate the IOL power. In the second stage, a selected lens with an extended focus range was implanted.
Results. The patient was examined and discharged on the second day after surgery with visual acuity OD 0.3 sph +0.25 cyl -0.75 ax150 = 0.5 and signs of moderate postoperative keratopathy. After 2 weeks, uncorrected visual acuity was as high as 0.9 and could not be further corrected by spectacle lenses. The patient was admitted for surgical treatment of the left eye cataract and also operated in two stages. The AL measured by the optical method in the aphakic state, was 0.31 mm less than that measured before the operation. Six months after PE with IOL implantation, distance visual acuity in both eyes was 1.2, the patient could use binocular vision for reading and PC work with no difficulty.
Conclusion. The double-step approach to IOL power calculation is justified in cases of strong decrease of crystalline lens transparency, especially in complicated cases, for example, in high myopia or “premium” lens implantation, in which patients are especially sensitive to correct target refraction, even though this approach requires a delayed IOL implantation, which should take place after a precise calculation of AL taken in the aphakic state.
About the Authors
A. N. KulikovRussian Federation
Aleksey N. Kulikov — Dr. of Med. Sci, professor, head of ophthalmology chair named after professor V.V. Volkov
6, Academician Lebedev St., St. Petersburg, 194044, Russia
E. V. Danilenko
Russian Federation
Ekaterina V. Danilenko — Cand. of Med. Sci., head of cataract surgery department, clinic of ophthalmology
6, Academician Lebedev St., St. Petersburg, 194044, Russia
N. V. Nevin
Russian Federation
Nikolay V. Nevin — ophthalmologist of cataract surgery department, clinic of ophthalmology
6, Academician Lebedev St., St. Petersburg, 194044, Russia
E. Yu. Kozhevnikov
Russian Federation
Evgeny Yu. Kozhevnikov — resident at ophthalmology chair named after professor V.V. Volkov
6, Academician Lebedev St., St. Petersburg, 194044, Russia
References
1. Tehrani M, Krummenauer F, Blom E, Dick HB. Evaluation of the practicality of optical biometry and applanation ultrasound in 253 eyes. J Cataract Refract Surg. 2003; 29: 741–6. doi: 10.1016/S0886-3350(02)01740-6
2. Fercher AF, Mengedoht K, Werner W. Eye-length measurement by interferometry with partially coherent light. Opt Lett. 1988 Mar 1; 13 (3): 186–8. doi: 10.1364/ol.13.000186
3. Rajan MS, Keilhorn I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations. Eye (Lond). 2002 Sep; 16 (5): 552–6. doi: 10.1038/sj.eye.6700157
4. Danilenko E.V. Sources of errors in calculation of intraocular lens power. Review of Literature. Oftalmokhirurgiya. 2012; 1: 85–8 (In Russ.).
5. Curtin BJ, Karlin DB. Axial length measurements and fundus changes of the myopic eye. I. The posterior fundus. Trans Am Ophthalmol Soc. 1970; 68: 312–34. PMID: 5524211
6. Melles RB, Holladay JT, Chang WJ. Accuracy of intraocular lens calculation formulas. Ophthalmology. 2018; 125 (2): 169–78. doi: 10.1016/j.ophtha.2017.08.027
7. Cooke DL, Cooke TL. Comparison of 9 intraocular lens power calculation formulas. J Cataract Refract Surg. 2016; 42: 1157–64. doi: 10.1016/j.jcrs.2016.06.029
8. Savini G, Taroni L, Hoffer KJ. Recent developments in intraocular lens power calculation methods — update 2020. Annals of Translational Medicine. 2020; 8 (22): 1553–62. doi: 10.21037/atm-20-2290
9. Malyugin B.E. State-of-the-art cataract surgery and intraocular optical correction. Vestnik oftal’mologii. 2014; 6: 80–8 (In Russ.).
10. Yu Y, Chen X, Hua H, et al. Comparative outcomes of femtosecond laserassisted cataract surgery and manual phacoemusification: a six-month follow-up. Clin Exp Ophthalmol. 2016; 44 (6): 472–80. doi: 10.1111/ceo.12695
11. Kulikov A.N., Danilenko E.V., Nevin N.V. Method for selecting the power of an implantable intraocular lens. Patent RF 2021131218; 2022 (In Russ.).
12. Hemmati HD, Gologorsky D, Pineda R 2nd. Intraoperative wavefront aberrometry in cataract surgery. Semin Ophthalmol. 2012 Sep-Nov; 27 (5–6): 100–6. doi: 10.3109/08820538.2012.708809
13. Kane JX, Chang DF. Intraocular lens power formulas, biometry, and intraoperative aberrometry: A review. Ophthalmology. 2021; 128 (11): e94-e114. doi: 10.1016/j.ophtha.2020.08.010
14. Kulikov A.N., Kokareva E.V., Dzilikhov A.A. Effective lens position. A review. Fyodorov journal of ophthalmic surgery. 2018; 1: 92–8. doi: 10.25276/0235-4160-2018-1-92-97
Review
For citations:
Kulikov A.N., Danilenko E.V., Nevin N.V., Kozhevnikov E.Yu. Double-step calculation of the intraocular lens power in lens opacities excluding the measurement of axial eye length by optical biometry. A clinical case. Russian Ophthalmological Journal. 2023;16(4):127-132. (In Russ.) https://doi.org/10.21516/2072-0076-2023-16-4-127-132